INFANT
INCUBATOR
The
D4S Baby
Incubator with Skin and Air control system offers
highest safety and optimum comfort for the new born.
Special
features,
-
The controller is based on Micro computerized system.
-
Large bright displays are easy to monitor and for viewing
from a distance.
-
Digital display system provided to measure the Infant
temperature, Air temperature and to set the required temperature
-
Feather touch keys provided for easy operation.
-
Bar graph LED system provided to indicate the heater output.
-
Fault indication and alarms for
• Baby probe failure
• Air probe failure
• High temperature
• Low temperature
• High temperature cut off
• Heater failure
• Fan failure
• Power failure
-
Baby probe / Air probe is made of highly sensitive sensor
and interchangeable which does not need field calibration,
the probe is made up of silicon sheathed cable, which
allows easy cleaning.
-
IV stand provided
-
Storage drawers provided
-
Large Baby tray provided
-
Trendelenburge / fowler system provided
-
Unit is powder coated
- Hygrometer
provided to indicate the relative humidity in the baby
compartment.
| Width |
: |
1040
mm |
| Depth |
: |
600
mm |
| Height |
:
|
1450
mm |
| Weight |
:
|
45
kgs Approx. |
| Baby
Bed Size |
:
|
550
x 390 mm |
| Tilt |
:
|
+
/ - 8o C |
| Recommended
Environment for operation |
: |
24o
C to 30o C. |
| Humidity |
: |
0
to 95 % R / H |
| Electrical |
:
|
230
V AC + / - 10 % 50 Hz. |
| Heater
Wattage |
:
|
250
Watts Max. |
| Over
all Wattage |
: |
450
Watts when heater, Fan is ON. |
INFANT
RADIANT WARMER
-
The
controller is based on Micro computerized system.
-
Servo and Manual Modes.
-
Self test system provided
-
Large bright displays are easy to monitor and for viewing
from a distance.
-
Double digital display system provided to measure the
Infant temperature and to set the required temperature
-
Feather touch keys provided for easy operation.
-
Bar graph LED system provided to indicate the heater output.
-
Fault indication and alarms for
• Baby probe failure
• High temperature
• Low temperature
• High temperature cut off
• Heater failure
• Power failure
-
Baby probe is made of highly sensitive sensor and interchangeable
which does not need field calibration, the probe is made
up of silicon sheathed cable, which allows easy cleaning.
-
Single element imported CE marked ceramic heater placed
in a parabolic reflector.
-
Detachable baby trolley provided with large baby tray
made up of stainless steel frame with three side acrylic
collapsible walls.
-
IV stand provided
-
Trendelenburge / fowler system provided
-
Unit is powder coated
-
Observation lamp provided
-
Monitor tray provided
| PHYSICAL
PARAMETERS : |
|
|
| |
|
|
| Length |
: |
700
mm |
| Width |
: |
430
mm |
| Working
level from the ground |
:
|
840
mm |
| Weight
Approx |
:
|
40
Kg |
| Castors |
:
|
75
mm |
| |
|
|
| |
|
|
| RECOMMENDED
ENVIORONMENTAL SPECIFICATIONS: |
| |
|
|
| Operating
temperature range: |
: |
20o
C to 33o C. |
| Storage
Temperature range |
:
|
-25o
C to 60o C |
| Operating
humidity range |
: |
0
– 100 % RH |
| |
|
|
| |
|
|
| SKIN
TEMPERATURE |
|
|
| |
|
|
| Range |
:
|
32o
C to 38o C |
| Accuracy |
:
|
+
/ - 0.2o C |
| Resolution |
:
|
0.1o
C |
| Probe
Interchangeability |
:
|
+
/ - 0.2o C |
| Infant
required temperature range |
:
|
32o
C to 38o C |
| Heater
output controllable from 0 to 100 % in manual mode (continuously
variable) |
| |
|
|
| |
|
|
| OBSERVATION
LAMP |
:
|
Halogen
based. |
| |
|
|
| |
|
|
| ELECTRICAL
SPECIFICATION |
|
|
| |
|
|
| Electrical
supply |
:
|
230
V AC + / - 10 % 50 Hz. |
| Power
consumption with heater at max output |
:
|
750
W. |
| |
|
|
RESUSCITATION
UNIT
PHOTOTHERAPY:
A halogen based phototherapy unit provided for Hyperbilirubinaemia
management. Three halogen lamps are fitted on either side
of the over head box. It is fitted in an angle position to
focus the radiation on the Infant for maximum effectiveness.
APGAR
TIMER:
The digital apgar timer is switched ON at the moment of birth
and an acoustic alarm will sound after 1, 5, 10, 15 and 20
minutes, reminding to examine the new born according to the
Apgar scheme.
X-RAY
CASSETTE TRAY:
The Neonatal Resuscitation unit heater box swivels away from
the baby bed to make room for X-ray unit. Convenient tray
assists you to achieve X-rays without disturbing the baby.
ELECTRICAL
SLOW SUCTION:
The slow suction unit is designed to work on latest diaphragm
technology and results in being maintenance free, noiseless
and provides high performance. The vacuum level can be adjusted
with the regulator. A vacuum gauge provided to monitor the
pressure.
RESUSCITATOR:
The D4S Resuscitator allows the resuscitation of distressed
newborns in delivery room or in case of emergency. Reanimation
is performed with positive pressure to the mask or endo tracheal
tube with manual control and free expiration. The minimum
and maximum pressure values can be preset and are clearly
indicated on the monometer. The Resuscitator operates when
connected to the air oxygen mixer or through any other flow
meter.
| PHYSICAL
PARAMETERS : |
|
|
| Mattress |
|
|
| Length |
: |
710
mm |
| Width |
: |
500
mm |
| Working
level from the ground |
:
|
900
mm |
| Weight
Approx |
:
|
45
Kg |
| Castors |
:
|
100
mm with 2 brake wheels and 2 non brake wheels. |
| |
|
|
| |
|
|
| RECOMMENDED
ENVIORONMENTAL SPECIFICATIONS: |
| |
|
|
| Operating
temperature range: |
: |
20o
C to 33o C. |
| Storage
Temperature range |
:
|
-25o
C to 60o C |
| Operating
humidity range |
: |
0
– 100 % RH |
| |
|
|
| |
|
|
| SKIN
TEMPERATURE |
|
|
| |
|
|
| Range |
:
|
32o
C to 38o C |
| Accuracy |
:
|
+
/ - 0.2o C |
| Resolution |
:
|
0.1o
C |
| Probe
Interchangeability |
:
|
+
/ - 0.2o C |
| Infant
required temperature range |
:
|
32o
C to 38o C |
| Heater
output controllable from 0 to 100 % in manual mode (continuously
variable) |
| |
|
|
| |
|
|
| PHOTOTHERAPY
UNIT |
|
|
| |
|
|
| Supply
to each lamp |
:
|
12
V, 50 W. |
| |
|
|
| |
|
|
| OBSERVATION
LAMP |
:
|
Halogen
based. |
| |
|
|
| |
|
|
| ELECTRICAL
SUCTION |
|
|
| |
|
|
Maximum
suction
|
:
|
400
mm of Hg. |
| Vacuum
gauge range |
:
|
0
-760 mm of Hg |
| |
|
|
| |
|
|
| ELECTRICAL
SPECIFICATION |
|
|
| |
|
|
| Electrical
supply |
:
|
230
V AC + / - 10 % 50 Hz. |
| Power
consumption with heater at max output |
:
|
750
W. |
| Power
consumption with heater, Phototherapy and Observation
Lamp ON |
:
|
1250
W. |
| |
|
|
| NOTE:
BEST RESULTS OBTAINED IN LOW CONVECTIVE ENVIRONMENT. |
PHOTOTHERAPY
UNIT:
The
D4S Phototherapy units are established for the treatment of
hyperbilirubinaemia, which greatly reduces the need to resort
to costly, relatively complex, exchange transfusion of blood.
The medical blue tubes give high output of light in the therapeutic
range, while the extended spectrum of the white light tubes
allow better observation of the Infant’s skin color
while in use.
The
Phototherapy units are mounted on a light weight mobile stand
with four castors. They are easy to use and have the following
features,
HEAD
TILT:
Top tray is tiltable and it can be locked at any desired angle.
This feature is of great help when used with radiant warmers.
HEIGHT
ADJUSTMENT:
The height adjustment facility enables the unit to be used
with a variety of Baby Incubators, cribs and costs.
| TECHNICAL
SPECIFICATIONS |
|
|
| |
|
|
| Stand
Type Phototherapy Unit |
|
|
| |
|
|
| Height |
: |
1565
mm |
| Width |
:
|
730
mm |
| Length
|
:
|
500
mm |
| Weight |
:
|
25
Kg |
| Castors |
:
|
50
mm |
| Finish |
: |
Powder
coated |
| |
|
|
| Power
Supply |
: |
230
V + / - 10 % 50 Hz |
| Power
Consumption |
:
|
20
W each lamp |
| Total
power requirement |
:
|
240
W |
| Main
fuse |
: |
1.5
A |
| |
|
|
| Trolley
Type Phototherapy Unit |
|
|
| |
|
|
| Height |
: |
1015
mm |
| Width |
:
|
730
mm |
| Length
|
:
|
525
mm |
| Weight |
:
|
20
Kg |
| Castors |
:
|
100
mm |
| Finish |
: |
Powder
coated |
| |
|
|
| Power
Supply |
: |
230
V + / - 10 % 50 Hz |
| Power
Consumption |
:
|
20
W each lamp |
| Total
power requirement |
:
|
160
W |
| Main
fuse |
: |
1.0
A |
To
get more details about Baby