FEATURES |
SPECIFICATION |
| |
|
|
|
| |
Construction |
Tabular,
Rigid, Complete Stainless Steel Sections or Electrostatically
Powder Coated Mild steel sections. Rigid stainless steel
top tray (size 597mm x 360 mm) Extended rear platform
for two 10 Ltr. water capacity cylinders. |
| |
|
|
|
| |
Mobility
|
Four
large diameter Anti static Castor wheels with double ball
bearing for anti-friction movement. Front Castors have
break. |
| |
|
|
|
| |
Colour
Coded Cylinder Yokes |
Gas
specific(Pin-indexed) Yokes with sliding clamping bars
for easy handling.
Two
each for oxygen and nitrous oxide (5 litre water capacity)
cylinders [Cylinders to be ordered separately) |
| |
|
|
|
| |
Colour
coded & Pressure Gauges |
Large
diameter (63 mm) two each for Oxygen and Nitrous oxide,
fitted at convenient angle for high visibility and clarity. |
| |
|
|
|
| |
Regulators |
Designed
for higher efficiency and reliability; two each for
Oxygen and Nitrous oxide.
Nitrous
oxide regulators are activated only when Oxygen regulators
are pressurised to minimise the risk of administering
100% nitrous oxide. |
| |
|
|
| |
OFWD
(Oxygen Failure Warning Device) |
Pneumatic
device which activates audio-visual alarm when oxygen
supply pressure falls to 205 kPa. |
| |
|
|
| |
Rotameters |
Long
(230 mm) rotating bobbling flowmeters calibrated in
double/ triple scale for accuracy and clarity in reading
with Luminous Back Plate which highlights rotating
Bobbin even in darkened O.T.
Oxygen
: 100 ml/min to 8 liters/min
Nitrous
Oxide : 200 ml/min to 12 liters/min |
| |
|
|
| |
Vaporisers |
One
Plenum type Ether Vaporiser fuly made in Brass & Copper
and Goldman Halothene Vaporiser supplied as standard.
Space for more vaporiser of choice is provided. |
| |
|
|
| |
Non-return
cum pressure Relief Valve |
Minimises
risk of backflow of gases; blows off when pressure exceeds
200 cm of water column. |
| |
|
|
| |
Breathing
Attachment |
Standard
magill's Circuit or Bains Circuit, Jackson Rees Circuit
with changeover unit (open circuit to closed circuit
& Vice versa).
Emergency
Oxygen flush at table top level. |
| |
|
|
| |
Accessories |
- Two built-in self sealing Schrader type oxygen outlets
(4.22 kg./cm2) For driving ventilators etc.
- One 60 psi Oxygen self sealing outlet at conventional
position with flowmeter & Humidifier Bottle.
- One Saline Stand
- B.P. Apparatus
- Coloured Plastic Bins
- Mox Oxygen Connection unit to attach Bulk cylinders
to Yokes.
|
| |
|
|
| |
Circle
Absorber |
Complete
Brass Body Circle Absorber with visible Inspiratory and
Expiratory Valve, Transperent Double chamber Sodalime
Cannister which can be easily reversed and fitted. A bypass
controlled through lever to divert gas flow from Sodalime
Cannister when not in use. A Circuit compramising- 2 corrugated
Hoses attached to Swivel 'y' Piece and a 2 Ltr. Bag with
hose. |
| |
|
|
| |
Dimentions |
Height |
156
cm |
| |
|
Depth |
66
cm |
| |
|
Width
|
80
cm |
| |
|
Weight
(Unpacked) |
70
kg |
| |
|
|
|
| |
|
|
|
| |
|
|
|
 |
D4S
MK III (ES) - Stainless Steel Trolley |
| D4S
MK III (EM) - M.S. Powder Coated Trolley |
|
Standard
Apparatus with sturdy built and aesthetic look having
all major features such as Rotameter Set, Plenum Type
Ether Vaporiser fully made in brass, Goldman Halothene
Vaporiser, Non-Return cum pressure relief valve, Outlet
with Emergency oxygen Flush at table top level, Mox
Oxygen connection unit, Magill's OR Bain Circuit, 60
psi Oxygen outlet port with Flowmeter and Humidifier
Bottle, with safety features such as Nitrous oxide Cut
-off system & Oxygen failure warning device and
standard spares.
Note
: Provision to attach circle absorber.
|
| |
|
|
|
| |
|
|
|
 |
D4S
Basic 'S2' Apparatus - Stainless steel trolley |
| D4S
Basic 'M2' Apparatus - M.S. Powder Coated Trolley |
A
compact, aesthetic and sturdy apparatus ensuring continuous
Oxygen supply arrangement with basic features like Rotameter
Set, Plenum Type Ether Vaporiser fully made in brass,
Goldman Halothene Vaporiser, Non-Return cum pressure relief
valve, Outlet with Emergency oxygen Flush at table top
level, Mox Oxygen connection unit, Magill's OR Bain Circuit,
60 psi Oxygen outlet port with Flowmeter and Humidifier
Bottle, with safety features such as Nitrous oxide Cut
-off system & Oxygen failure warning device and standard
spares. Note
:
-
Provision to keep two 10 Ltr. water capacity cylinders.
- Provision
to attach circle absorber.
|
| |
|
|
|
| |
|
|
|
 |
D4S
Basic 'S1' Apparatus - Stainless steel trolley |
| D4S
Basic 'M1' Apparatus - M.S. Powder Coated Trolley |
A
compact, aesthetic and sturdy apparatus with basic anaesthesia
components featuring Rotameter Set, Plenum Type Ether
Vaporiser fully made in brass, Goldman Halothene Vaporiser,
Non-Return cum pressure relief valve, Outlet with Emergency
oxygen Flush at table top level, Mox Oxygen connection
unit, Magill's OR Bain Circuit, 60 psi Oxygen outlet port
with Flowmeter and Humidifier Bottle, with safety features
such as Nitrous oxide Cut -off system & Oxygen failure
warning device and standard spares. Note
:
-
Provision to keep two 10 Ltr. water capacity cylinders.
- Provision
to attach circle absorber.
|
| |
|
|
|
| |
|
|
|
 |
D4S
Economy 'S' Apparatus- Stainless steel trolley |
| D4S
Economy 'M' Apparatus- M.S. Powder coated trolley |
|
A
most economical compact apparatus with basic anaesthesia
components featuring - Rotameter Set, Plenum Type Ether
Vaporiser fully made in brass, Goldman Halothene Vaporiser,
Outlet with Emergency oxygen and Relief Valve at Vaporiser
level , Mox Oxygen connection unit, Magill's OR Bain
Circuit with safety features such as Nitrous oxide Cut
-off system & Oxygen failure warning device and
standard spares.
Thus
making it most cost effective apparatus without compromising
safety features.
Note
:
-
Provision to keep two 10 Ltr. water capacity cylinders.
- Provision
to attach circle absorber.
|
| |
|
|
|
| |
|
|
|
| D4S
Portable Anaesthesia Apparatus |
 |
| |
|
| Simple,
safe, reliable, complete stainless steel compact structure
making it a good portable, easy to transport model with
features as Standard Rotameter, Goldman Halothene Vaporiser,
Outlet unit with Relief Valve and Emergency Oxygen flush,
Magill's OR Bain Circuit, Seperate Oxygen and Nitrous
oxide Preset regulators with Coloured Coded Tubings and
diffrent end fittings to avoid accidental interchangeability,
Nitrous Cut -off system & standard spares. |
| |
|
|
 |
| |
OPTIONAL
: |
|
|
| |
|
|
|
| |
1.
Hypoxia Guard with ORC facility at Rotameter |
| |
Compact
Redction Gear Mechanism (No Link Chain) which ensures
a minimum of 25% (± 5%) Oxygen Concentration in
gas mixture which can be raised anywhere in the range
of 25% to 100% |
| |
|
|
|
| |
|
|
|
 |
|
|
| 2.
Mainstream Oxygen Monitor |
In-built
Oxygen Concentration Monitor measures the oxygen Concentration
directly in the breathing system at patient end with Audio-Visual
Hypoxia Alarm.
|
| |
|
|
|
| |
|
|
|
| |
|
|
 |
| |
3.
Medical Gas Delivery System |
| |
WE
undertake installation of Central Pipeline for Medical
Gases, Vacuum and compressed Air to wards, O.T.'s and
I.C.U.'s which is becoming essential requirement in Hospital
and Nursing Home. |
| |
 |
|
|
| |
|
|
|